Conductivity Analyzers

ABB conductivity analyzers are designed to meet the most exacting requirements of our valued customers. The extensive range covers 2 and 4-wire analyzers and single or dual-input devices.

Our hazardous area certifications embrace FM, CSA and CENELEC ATEX intrinsic safety. Instruments are available with HART, Profibus PA or DP, and Foundation Fieldbus communications capabilities.

2-Electrode Contacting Cells - Range 0 to 10,000 µS/cm
High accuracy technique for trace conductivities from ultrapure water to industrial rinse waters. An alternating current is passed between two electrodes in contact with a solution. The transmitter converts the resultant potential difference into conductivity, concentration or specific resistance.

4-Electrode Contacting Cells - Range 0 to 2,000 mS/cm
Auto-compensating technology for high level conductivity and concentration measurements in high-coating processes. Provides an alarm when cleaning is required.

Electrodeless sensors - Range 0 to 2,000 mS/cm
Non-contact technique providing immunity to polarisation, oily fouling and non-conducting coatings until gross fouling occurs. Measurement of the magnitude of induced current into a closed loop of solution using two embedded toroids.

United States Pharmacopœia <645>
Traceable calibration instruments for WFI and PW processes in the Pharmaceutical Industry.
W
W
2-Electrode Contacting Cells - Range 0 to 10,000 µS/cm (depending on cell constant)

AX410
AX411
AX416
Input 1
µS/cm
µS/cm
µS/cm
Input 2
-
µS/cm
pH
Application
4-wire AC &
24V DC
4-wire AC &
24V DC
4-wire AC &
24V DC
Communications
4 to 20mA, Profibus DP
4 to 20mA, Profibus DP
4 to 20mA, Profibus DP
Mounting
Panel, Wall, Pipe
Panel, Wall, Pipe
Panel, Wall, Pipe
Hazardous Area
No
No
No
W
ACA592*
TB82TE*
TB84TE*
Input 1
µS/cm
µS/cm
µS/cm
Application
2-wire 24V DC
2-wire 24V DC
4-wire AC
Communications
HART
4 to 20mA, HART, FF,
Profibus PA
4 to 20mA
Mounting
Wall, Pipe
Panel, Wall, Pipe
Panel, Wall, Pipe
Hazardous Area
FM, CSA, ATEX
FM, CSA, ATEX
FM, CSA
*0 to 20,000 µS/cm
W
W
4-Electrode Contacting Cells - Range 0 to 2,000 mS/cm (depending on cell constant)
AX430
AX436
ACA592
TB82EC
TB84EC
Input 1
µS/cm
µS/cm
mS/cm
mS/cm
mS/cm
Input 2
-
pH
-
-
-
Application
4-wire AC &
24V DC
4-wire AC &
24V DC
2-wire 24V DC
2-wire 24V DC
4-wire AC
Communications
4 to 20mA, Profibus
4 to 20mA, Profibus
HART
4 to 20mA, HART,
FF, Profibus PA
4 to 20mA
Mounting
Panel, Wall, Pipe
Panel, Wall, Pipe
Wall, Pipe
Panel, Wall, Pipe
Panel, Wall, Pipe
Hazardous Area
No
No
FM, CSA, ATEX
FM, CSA, ATEX
FM, CSA
W
W
Electrodeless sensors - Range 0 to 2,000 mS/cm
ACA592
TB82TC
TB84TC
Input 1
mS/cm
mS/cm
mS/cm
Input 2
-
-
-
Application
2-wire 24V DC
2-wire 24V DC
4-wire AC
Communications
HART
4 to 20mA, HART,
FF, Profibus PA
4 to 20mA
Mounting
Wall, Pipe
Panel, Wall, Pipe
Panel, Wall, Pipe
Hazardous Area
FM, CSA, ATEX
FM, CSA, ATEX
FM, CSA
W
W
United States Pharmacopœia <645>
AX450
AX455
Input 1
µS/cm
µS/cm
Input 2
-
µS/cm
Application
4-wire AC
4-wire AC
Communications
4 to 20mA, Profibus DP
4 to 20mA, Profibus DP
Mounting
Panel, Wall, Pipe
Panel, Wall, Pipe
Hazardous Area
No
No

    •   Cancel